Alarms, messages, and troubleshooting
9-17
Vent Inoperative
1005
Internal Tempera-
ture High
Technical failure of the
motor PCBA. The venti-
lator is in the ventilator
inoperative state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(7)
No No No
Vent Inoperative
1006
Data Acquisition
PCBA ADC Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(8)
No No No
Vent Inoperative
1007
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(9)
No No No
Vent Inoperative
1008
Machine and
Proximal Pres-
sure Sensors
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(10)
No No No
Vent Inoperative
1009
Pressure Regula-
tion High
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(11)
No No No
Vent Inoperative
100A
Data Acquisition
PCBA ADC Refer-
ence Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(12)
No No No
Vent Inoperative
100B
Watchdog Test
Failed
Technical failure. The
ventilator is in the ven-
tilator inoperative
state.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
High
(13)
No No No
Table 9-5: Vent Inoperative alarm messages: summary and troubleshooting (continued)
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able