General information
3-11
CAUTION: Avoid allowing the ventilator battery to become completely
discharged. Otherwise, the battery may become over-discharged and
require long recharge times of up to 16 hours or more. The over-
discharged condition may permanently damage the battery so that it
is unable to recharge. To prevent the occurrence of a non-recoverable
over-discharged battery, always keep the ventilator connected to an
AC outlet.
Figure 3-6: Power indicators