: Pressing the key triggers a single pulse of the preset increase duration T.
(4) Now use the Data Selector <2> to click on the symbol to return to the start menu.
The manual release key symbol is now displayed next to the pulse symbol .
Important
If you use the manual release key, you can only turn the intensity up in both channels if
you press the manual release key (security blocking)!
(5) With the Data Selector <2> click on the symbol in channel II. The following window opens:
Selection window for channel II
(6) First use the Data Selector <2> to click on the R symbol, turn to the left until the desired symbol
(simultaneous or alternating stimulation) appears and press to confirm.
(7) Now use the Data Selector <2> to click on the symbol to return to the start menu.
The respective symbol for simultaneous or alternating simulation now appears next to the pulse sym-
bol of channel II.
How to Control the Pulse Currents Manually with Preselected Pulse Duration
(1) In the Display <1>, select a suitable current mode e.g. KOTS.
(2) With the Data Selector <2> click on the symbol in channel I. The following window opens:
Selection window for channel I
(3) First use the Data Selector <2> to click on the symbol R, turn to the right until the until the manual
release key symbol appears and press to confirm.
(4) Now use the Data Selector <2> to click on the symbol to return to the start menu.
The manual release key symbol is now displayed next to the pulse symbol .
Important
If you use the manual release key, you can only turn the intensity up in both channels if
you press the manual release key (security blocking)!
(5) Now use the Data Selector <2> to click on the symbol T, set the pulse duration by turning and con-
firm by pressing it.
(6) With the Data Selector <2> click on the symbol in channel II. The following window opens:
3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG