Specification Value
Collet clamping range 3 mm (0.1 inches)
Maximum speed 100 000 rpm
Weight ~ 2.3 kg (~ 5.1 lbs)
Connection values
Specification Value
Voltage 110 / 230 V
Nominal current Max. 1.5 A
Frequency 50 / 60 Hz
Nominal power Max. 500 W
Required air pressure 3 to 4 bar
Required air flow rate 50 l/min
Main fuse 6 A
Environmental requirements
The individual components of the machine possess varying expansion
coefficients. Imprecision during processing can only be ruled out with an
ambient temperature from +18 to +25°C. If your ambient temperatures are
not within this range, we recommend an air conditioner. Contactyour local
Planmeca dealer for this purpose.
Requirement Specification Value
Operation Temperature range +18 to +25 °C
+64 to +77 °F
Max. humidity 60 %
Storage Temperature range +10 to +50 °C
+50 to +122 °F
Max. humidity 80 %
Transport Temperature range -10 to +55 °C
+14 to +131 °F
Max. humidity 80 %
Requirements for compressed air connection
The compressed air connected must expressly comply with the following
conditions.
Specification Value
Air pressure [P] 3 to 4 bar
Connection Plug-in nipple NW 7.2 mm (quick-connection)
Air purity
Specification according to ISO 8573-1, Compressed air for general use, part
1: Contaminants and quality classes.