Chapter A - GENERAL & TECHNICAL DATA
A-4 Planmeca Proline XC X-ray unit
TECHNICAL SPECIFICATIONS
Technical Manual
Mechanical data
Weight Panoramic:108 kg / 238 lbs
Pan/ceph 126 kg / 278 lbs
Dimensions (HxDxW) 2200x930x890 mm, 86.6x36.6x35 inch
Color RAL 9002
Environmental requirements
Ambient temperature Operating +5°C to +40°C
Storage -10°C to +50°C (film X-ray unit)
0°C to +50°C (Dimax3 digital X-ray system)
Humidity 25% - 75%
Dimax3 digital X-ray system
CCD pixel size 33 μm
Image pixel size 130 μm normal resolution
99 μm enhanced resolution/
66 μm high resolution
Image size 2144x1032 pixels normal resolution
2860x1376 pixels enhanced resolution
4288x2064 pixels high resolution
Sensor active area size Pan 9 x 136 mm
Ceph 9 x 230 mm
Image field Panoramic 14 x 30 cm (5.5 x 12 in.)
Cephalometric 22.5 x 30 cm (9 x 12 in.)
External mains fuse recommendation
The recommendation for the external mains fuses are:
• units with 100 - 132 V~ voltage setting: 16A, time lag
• units with 180 - 240 V~ voltage setting: 8A, time lag
No other equipment should be connected to the same fused mains line as the X-ray unit. In
some countries an additional external fault current guard is also required.
Manufacturer
PLANMECA Oy, Asentajankatu 6, FIN-00880 Helsinki, FINLAND
phone: +358-20-7795 500