Chapter A - GENERAL & TECHNICAL DATA
Planmeca Proline XC X-ray unit A-3
TECHNICAL SPECIFICATIONS
Technical Manual
3 TECHNICAL SPECIFICATIONS
X-ray tube Toshiba D-052SB
Focal spot size 0.5 x 0.5mm
according to IEC 60336
Target angle 5°
Total filtration min. 2.5 mm Al
Generator Constant potential, microprocessor controlled,
operating frequency 80 kHz
Anode voltage 60-80 kV ± 5%
Anode current 4-12 mA ±(5% + 0.5mA)
Exposure time Panoramic: 2.5- 18 s as indicated ±10%
Cephalometric: 17- 23 s ±10%
Cooling period Automatically controlled
Film size (film X-ray unit) Panoramic:15 x 30 cm
Cephalometric: 18 x 24 cm and 8” x 10”
The unit equipped with ceph film size 18 cm x 24 cm
X-ray beam size on film
in ceph mode (H x W) 230 mm x 170 mm (primary collimators 4 and 5)
170 mm x 230 mm (primary collimator 6)
The unit equipped with ceph film size 8 in. x 10 in.
X-ray beam size on film
in ceph mode (H x W) 245 mm x 190 mm (primary collimators 4 and 5)
190 mm x 245 mm (primary collimator 6)
Cassette (film X-ray unit) Flat
SID Panoramic: 480 mm (19”)
Cephalometric: 163 - 170 cm (64” - 67”)
Magnification Panoramic: constant 1.2
Cephalometric: 1.08 - 1.13
Line voltage/ 100-132 / 50 or 60 Hz
Mains frequency 180 - 240 V~ / 50 Hz
Apparent resistance 0.3 ohms 100 V~ / 0.8 ohms 230 V~
Regulation Automatic ± 10%
Line current max. 8A at 230V~, 15A at 100V~
Fusing 8FF 500 VAC at 180 - 240 V~
16FF 500 VAC at 100 -132 V~
Maximum continuous
heat dissipation < 250W
Electrical classification
Class I
Type B