Patient supports As shown in section Patient supports in User
Manuals
Patient handles
Generator (according to IEC 60601‐2‐7: 1998)
Resonant-mode, DSP-controlled, 80 ‐ 160 kHz
X‐ray tube
Toshiba D-059SBR or SXR 130-10‐0.5 SC
Focal spot size (according to IEC 60336: 2005)
0.5 x 0.5 mm
Filtration
3D Total 2.5 mm Al + 0.2 mm / 0.5 mm Cu
Pan (SmartPan) / ProCeph Total 2.5 mm Al
Tube housing front cover quality equivalent
filtration (not included in the specified total
filtration)
0.3 mm Al @ 70 kV / HVL 2.6 mm Al
Anode voltage
3D 80 ‐ 120 kV ±5%
Pan (SmartPan) 60 ‐ 84 kV ±5%
ProCeph 60 ‐ 84 kV ±5%
Anode current
3D Toshiba D-059SBR: 1‐12.5 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
Pan (SmartPan) Toshiba D-059SBR: 1‐14 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
ProCeph Toshiba D-059SBR: 14 mA ±10%
SXR 130-10‐0.5 SC: 16 mA ±10%
mAs range
min. / max. as indicated ±(10% + 0.2 mAs)
Dose range and accuracy
Dose range min. / max. as indicated on system
user interface.
Accuracy of dosimetric indication (DAP, CTDI):
±40%
Linearity of radiation output
< 0.1
Exposure time
3D Pulsed, effective 1.5 ‐ 36 s as indicated ±10%
Pan (SmartPan) 2.5 – 15.6 s as indicated ±10%
ProCeph 0.1 – 1.6 s as indicated ±10%
SID
3D / Pan (SmartPan) 700 mm
Ceph 1700 mm (66.9 in.)
1 General and technical data
Technical manual Planmeca Viso 17