CAUTION
Use protective eye-wear when changing the battery.
CAUTION
The maximum temperature of the tube head is limited to 60° C (140° F). The
rubber bellows inside the tube head keep the pressure constant within the
allowed temperature range.
CAUTION
EMC requirements have to be considered, and the equipment must be
installed and put into service according to the specific EMC information
provided in section "EMC information" on page 27.
CAUTION
Portable and mobile RF communications equipment can affect the X-ray
unit. For minimum distance between portable and mobile RF
communications equipment and the X-ray unit, see section "EMC
information" on page 27.
CAUTION
ELECTRICAL SAFETY RULES
Always replace the fuses with ones of the same type and rating. Otherwise
patient, operator or equipment safety cannot be guaranteed.
The circuit boards can be damaged due to static discharges and requires
careful handling.
CAUTION
GENERAL SAFETY RULES
The unit must be serviced only by qualified personnel, trained by
PLANMECA. Repairs and parts replaced by unqualified personnel carry no
warranty.
Periodical maintenance as described in this manual must be performed on a
regular basis, to ensure the safety and image quality of the unit.
Some procedures described in the unit could be dangerous, if not followed
as stated.
NOTE
The display values shown in this manual are only examples and should not
be interpreted as recommended values unless otherwise stated.
1.3 Symbols on product labels
Fulfils the requirements of Directive 93/42/EEC.
SGS listing marking according to US and Canadian standards (ANSI/AAMI
ES60601-1 and CAN/CSA C22.2 No. 60601- 1).
Date of manufacture (Standard ISO 7000).
1 General and technical data
Technical manual Planmeca Viso 3