E705 Tube filament
definition
Tubehead filament calibration
failure (discontinuous tube
current).
The current in the tube head is
not high enough. Perform the
preheat calibration again.
E706 Tubehead filament calibration
failure (mA too low).
The current in the tube head is
not raising on the voltage
area. Perform the preheat
calibration again. Check the
functionality of the tube head.
E710 Primary collimator Primary y-collimator opening is too
big.
Adjust the lower blade
upwards or the upper blade
downwards.
E711 3D Primary x-collimator opening is
too wide.
Adjust the left blade to the
right or the right blade to the
left.
E712 Collimator calibration failed.
3.5 System conflicts (E8xx)
The following table lists the system conflict errors.
Error
code
Explanation Comments
E814 Missing internal
parts
Missing internal parts, CAM not
found.
Check the functionality of the
CAM PCB.
E815 Missing internal parts, TUBE not
found.
E818 Missing internal parts, Ethernet
module not found.
Check the functionality of the
Ethernet PCB and that it is
attached properly on the CPU.
E821 Software
compatibility
CPU FPGA version not compatible
with the main software.
E822 CAM FPGA version not
compatible with main software.
Check the software version of
the Viso X-ray unit.
Error
code
Explanation Comments
E835 Movement prevented due to
missing jumper J8 on patient
position control PCB or broken
patient positioning cable.
E836 Wrong C-arm PCB version for
120kV device.
E841 Missing MCM Missing MCM module 1.
E842 Missing MCM module 2.
E843 Missing MCM module 3.
E844 Missing MCM module 4.
E845 Missing MCM module 5.
E846 Missing MCM module 6.
E847 Missing MCM module 7.
E848 Missing MCM module 8.
3 Error messages
46 Planmeca Viso Technical manual