4.3.2 Lubrication
Visually check for the amount of lubricant in the linear actuator and threaded
rods of the linear actuator and collimator.
If necessary, lubricate the parts with Shell Gadus S3 A1300C 2 grease, or a
grease with the same specifications as Shell Gadus S3.
After lubrication, wipe away the excess lubricant.
4.4 Software update
Perform software update according to the instructions provided by the
software update Wizard.
4.5 X-ray measurements and exposure values
There are two ways of doing most of the following tests:
• From the radiation beam
4 Annual maintenance
Technical manual Planmeca Viso 53