E401 Tube head offsets Tube head KVPOS offset failure,
short- circuited.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E402 Tube head KVPOS offset failure, out
of bounds.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E403 Tube head offset KVNEG failure,
short- circuited.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E404 Tube head KVNEG offset failure, out
of bounds.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E405 Tube head MAPOS offset failure,
short- circuited.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E406 Tube head MAPOS offset failure, out
of bounds.
Exposure not possible.
Check the cabling between
the FBK and the CAM
PCB. Check also the
functionality of the FBK
PCB.
E407 Tube head MANEG offset failure,
short- circuited.
Exposure not possible.
Check the functionality of
the FBK PCB.
E408 Tube head MANEG offset failure, out
of bounds.
Exposure not possible.
Check the functionality of
the FBK PCB.
E409 Tube head filament offset failure, out
of bounds.
Exposure not possible.
Check the functionality of
the FBK PCB.
Error
code
Explanation Comments
3 Error messages
Technical manual Planmeca Viso 43