7-2
7 Technical Specifications
Housing
Calibration
7.2.2 Signal Inputs and Outputs
REF 1 to REF 8
TRIG IN, AUX IN
SIGNAL 1 to SIGNAL 4
Protection rating: IP20
Dimensions (w x h x d): Without angle brackets
450mm x 150mm x 380mm (19", 84HP /3U)
With angle brackets
485mm x 150mm x 380mm (19", 84HP /3U)
Weight: 10 kg
Recommended calibration
interval: 2 years
Input voltage range: ±1 V and 10V
Input impedance: 1MΩ, in parallel with 100pF
Input coupling: AC/ DC, adjustable in the software
AC − 3 dB cutoff frequency: 0.05Hz
Overvoltage protection: ±30 V (against damage)
IEPE mode
Sensor supply: nom. 4mA/ 24V, IEPE compatible
Time constant: 3.3s
Compatibility: TTL/ CMOS
Overvoltage protection: ±30 V (against damage)
Input resistance: 10kΩ
Output voltage swing: max. ±10V at high-impedance load
max. ±5V into 50Ω
Output current: max. ±100mA
Output impedance: 50Ω
Short-circuit protection: permanently short-circuit proof