EasyManua.ls Logo

ProMinent DULCOMETER D1Cc - Reduced; Complete Operating Menu; Setting the Reading

ProMinent DULCOMETER D1Cc
155 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9.3 Reduced / Complete Operating Menu
The DULCOMETER
®
D1Cb / D1Cc allows settings to be made in two different compre‐
hensive menus (reduced / complete). All of the parameters of the controller are preset
and can be changed in the complete operating menu.
The controller is delivered with a restricted operating menu. If adjustments are neces‐
sary, the parameters can be changed by switching to the complete operating menu.
general setting
information
Identcode:
D1CB xxxxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxxx
alarm relay
access c.:
5000
operating menu
english
reduced
A0088_GB
operating menu
Calib timer
wash timer
_
_
Fig. 49: Reduced / complete changeover
9.4 Setting the Reading
WARNING!
Incorrect metering due to incorrect measured variable
Possible consequence: Fatal or serious injuries
The measured variable/measuring range of the sensor/measuring device is
essential for the measured variable/measuring range!
If the measured variable/measuring range is changed, the sensor must be recali‐
brated
If the measured variable/measuring range is changed, the set points and limit
values will be changed to the associated starting values!
If the measured variable/measuring range is changed, the settings in all of the
menus must be checked
Standard signal input mA
measured value
setting ?
measured value
range adjustment
20 mA
=
2
.00 ppm
measured value
range adjustment
4 m
A
= 0.00 ppm
A0019_GB
Fig. 50: Reading
Measured Variables and Operating Menus for the Standard Signal General
104

Table of Contents

Related product manuals