10.6 General settings
general setting
information
identcode
D1CBxxxxxxxxxx
Vxxxxxxxxxxx
Srnr: xxxxxxxxxx
alarm relay
Calib timer
access c.:
5000
operating menu
-
English
-
reduced
A0016_GB
pause
operating menu
Fig. 71: General settings
The functions described below can be selected in this menu.
10.6.1 Setting the Measured Variable/Measuring Range
WARNING!
Incorrect metering due to incorrect metering range
Possible consequence: Fatal or very serious injuries
– If the assignment of the measuring range is modified, the settings must be
checked in all menus
–
If the assignment of the measuring range is changed, the sensor or the meas‐
uring device must be recalibrated
– The measuring range of the sensor or the measuring device is essential for the
measuring range!
general setting
Information
Identcode:
Vxxxxxxxxxxxx
D1CBxxxxxxxxxxx
Srnr: xxxxxxxxxx
add. functions
Software version
measure change
wash timer chlorit
Br
O3
O2
effective range
A0032
0...2.00ppm
Fig. 72: Measuring range
In the "Change measured variable" setting range, it is possible to select the measured
variable required for the respective process and suitable for the sensor or measuring
device. Depending on the measured variable, different setting windows must be pro‐
cessed, which are provided by the software of the DULCOMETER
®
D1Cb / D1Cc. The
values are set, selected and confirmed using keys
, , and .
Operating Menus Independent of Measured Variables