8.6 Temperature correction value for pH and fluoride sensors
CAUTION!
Influence of temperature on the pH or fluoride measurement
Possible consequence: Slight or minor injuries. Material damage.
Temperature changes in the sample water lead to a change in the slope of the cali‐
bration lines (pH, fluoride) and to a displacement of the zero point with pH sensors or
the standard potential E
S
for fluoride sensors.
Measure to be taken to avoid this danger:
–
The pH or fluoride measurement should only be carried out in the
[Temperature Correction Value automatic]
setting
– The DULCOMETER
®
D1Cb / D1Cc then automatically compensates for both
effects when connecting a temperature sensor (Pt 100/Pt 1000)
only with correction value
correction value
setting ?
correcting value
temperature
automatic
A0020_GB
Fig. 47: Temperature correction value for pH and fluoride sensors
Temperature correction value for pH and fluoride sensors in line with the identity code:
Possible values
As per identity code Increment Lower value Upper value
0 off
2 off
Manual
Automatic
4 off
Manual
Measured Variables and Operating Menus for Potentiometric Sensors