EasyManua.ls Logo

ProMinent DULCOMETER D1Cc - Page 57

ProMinent DULCOMETER D1Cc
155 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DULCOMETER
®
D1Cb / D1Cc software upgrade
D1U
b
Software default setting
V Software preset
Default setting - measured variable
0 Universal
A Peracetic acid
B Bromine
C Chlorine
D Chlorine dioxide
F Fluoride
H Hydrogen peroxide
I Chlorite
P pH
R ORP
S 0/4-20 mA standard signal general
X Oxygen
Z Ozone
L Conductivity
Connection of the measured variable
1* Standard signal 0/4-20 mA, all measured variables
5 mV input for pH/redox via guard terminal
Correction variable
0 none
2* Temperature Pt100/PT1000 (for pH and conductivity)
4* Manual temperature input (for pH and conductivity)
* = chargeable option