EasyManua.ls Logo

Prunus Boaray 5000C - Page 61

Default Icon
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
3. Other settings
Touch Other settingsto access the other settings screen. It shown as below.
Figure 5-11 Other settings screen view
1) Atomize
Touch the option frame and select Disable, 10 minutes, 20 minutesor 30 minutes in
the drop-down menu.
When O
2
supply is connected to the system and a nebulizer is attached to the ventilator,
and the minimum flow is more than 10 L/min, after the nebulization button is pressed, the
ventilator will automatically supply 7L/min (±0.5L/min) gas to the patient in inspiration
phase, and the nebulization time can be selected in 10min, 20min and 30min (in system
setting screen). You may end the nebulization period early by pushing the nebulization
button again.
Note: When connecting nebulizer, it need install bacterium filter onto expiration end.
2) Percent of the leak
Touch the option frame and select the percent of the leak (Range:20~80%).
When the system occurs leak, for example: ( V
TI
- V
TE
/V
TI
> set leak percent, System
leakage!!! Please check the breathing tubes and expiratory valve. alarm is activated.
Default setting: 20%.
3) Waveform Selection
Waveform type can be set in VCV mode, SIMV+VCV mode and SPONT mode(Backup
ventilation: VCV).
Touch theWaveform Selectionoption frame to select Square or Degressive in

Table of Contents

Related product manuals