• Use the scale keyboard a starting correction value and confirm
it by pressing
.
Notice:
1. The starting correction value is estimated experimentally in at least 2
subsequent cycles of dosing;
2. In correction mode <CONSTANT>, the start correction is always the
correction value, but in correction mode <AUTO> the declared start
value is used only in the first dosing cycle. Then it can be modified.
Return to weighing:
See 11.3.
24.8.6.3. Maximum correctional value
The introduced maximum value in parameter <P7.4.6.3 MAX
CORRECTION>for corrections is intended to limit the range of
operation for automatic calculation of correctional value.
Procedure:
• Enter <P7 WORK MODES> and then:
• Use a keyboard to enter a maximum value for the process and
confirm it by pressing
.