27.7. RS485 led out via RS 1D gland
A version with the RS485 interface led out via a gland (in the standard
solution RS485 is present in a socket). A 3m cable is led out through
the gland.
27.7.1. Installing inside the PUE C41H housing
1. Unplug the scale from mains;
2. Unscrew and take off the back wall of the housing;
3. Unplug one of the unused glands and led out the PT0012
cable through it;
4. Connect the PT0012 cable to the J20 connector (ARK type)
on the main board of PUE C41H: