Connecting 4-wire tensometers
E - tensometer cable shield
REF+ - solder jumper JP1
REF- - solder jumper JP2
IN+ - “OUTPUT+” from tensometer
IN- - “OUTPUT-“ from tensometer
+5V - “INPUT+” from tensometer
AGND - “INPUT-“ from tensometer
The rules of connecting shields from tensometer cable
For assuring appropriate operation use the description below to connect
the shield of the tensometer properly. In both cases (6- and 4-wire cables)
the same way rule are valid: