Browser-Based User Interface
R&S
®
NRPxxS(N)
50User Manual 1177.5079.02 ─ 10
A value of 0.0 corresponds to an ideal matched source. A value of 1.0 corresponds to
total reflection.
Remote command:
[SENSe<Sensor>:]SGAMma:MAGNitude on page 114
Phase ← Γ Correction
Sets the phase angle of the complex reflection coefficient of the source, Γ
source
.
Remote command:
[SENSe<Sensor>:]SGAMma:PHASe on page 115
Zero Calibration
Performs zeroing using the signal at the sensor input. See Chapter 9.9, "Calibrating/
Zeroing the Power Sensor ", on page 125.
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
Remote command:
CALibration<Channel>:ZERO:AUTO on page 126
Diagnostics
Triggers a selftest of the sensor.
Note:
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps Offset Voltages and/or Noise Voltages.
The results of the selftest are shown in a dialog that is displayed after the test comple-
tion.
Settings

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals