Remote Control Commands
R&S
®
NRPxxS(N)
89User Manual 1177.5079.02 ─ 10
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxS(N) sets the last used
length.
Example for REAL,32 format:
#6768000....<binary float values>....
Example for REAL,64 format:
#71536000....<binary float values>....
<data[,length]> <ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for ASCii,4 format:
1.2938e-06, -4.7269e-11, ...
*RST: ASCii,0
9.6.3 Retrieving Results
After performing the measurement, you can query the measurement results with a
command from the FETCh subsystem.
Remote commands:
[SENSe<Sensor>:]FUNCtion.............................................................................................89
FETCh<Sensor>:ARRay[:POWer][:AVG]?.......................................................................... 90
FETCh<Sensor>[:SCALar][:POWer][:AVG]?........................................................................90
FETCh<Sensor>[:SCALar][:POWer]:BURSt?......................................................................91
FETCh<Sensor>[:SCALar][:POWer]:TSLot?....................................................................... 91
CALCulate:FEED.............................................................................................................91
[SENSe<Sensor>:]AUXiliary..............................................................................................92
[SENSe<Sensor>:]FUNCtion <function>
Sets the measurement mode.
Parameters:
<function> "POWer:AVG"
Continuous Average
After a trigger event, the power is integrated over a time interval
(averaging) set with [SENSe<Sensor>:][POWer:][AVG:
]APERture.
Configuring and Retrieving Results