Remote Control Commands
R&S
®
NRPxxS(N)
94User Manual 1177.5079.02 ─ 10
Using [SENSe<Sensor>:][POWer:][AVG:]FAST ON, you can accelerate the mea-
surement as follows:
Chopper is deactivated.
Average count is set to 1, no matter which average count you have set.
Thus, the overall measurement time is only defined by the aperture time, and the mea-
surement time for a fast measurement is calculated as follows:
MT = APER
A programming example is given in Chapter 10.2, "Performing the Fastest Measure-
ment in Continuous Average Mode", on page 144.
Remote commands:
[SENSe<Sensor>:][POWer:][AVG:]APERture......................................................................94
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar...............................................................94
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?............................................................94
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?.............................................................. 95
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE.................................................................95
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe...............................................................95
[SENSe<Sensor>:][POWer:][AVG:]FAST............................................................................ 95
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe..........................................................96
[SENSe<Sensor>:][POWer:][AVG:]APERture <integration_time>
Sets the duration of the sampling window in the continuous average mode. During this
time interval, the average signal power is measured.
Parameters:
<integration_time> Range: 10.0e-6 to 2.00
*RST: 0.02
Default unit: Seconds
Example:
APER 0.02
Manual operation: See "Aperture Time" on page 45
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer in continuous average mode.
Example:
BUFF:CLE
Usage: Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Queries the number of results that are currently stored in the result buffer. Available in
continuous average mode.
Example:
BUFF:COUN?
Usage: Query only
Configuring the Measurement Modes

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals