Remote Control Commands
R&S
®
NRPxxS(N)
95User Manual 1177.5079.02 ─ 10
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast the FETCh? command returns a result only if the buffer is full.
Usage: Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE <count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using [SENSe<Sensor>:][POWer:][AVG:]BUFFer:
STATe.
Parameters:
<count> Range: 1 to 8192
*RST: 1
Example:
BUFF:SIZE 1
See Chapter 10.3, "Performing a Buffered Continuous Average
Measurement", on page 146.
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe <state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer with [SENSe<Sensor>:][POWer:][AVG:
]BUFFer:SIZE.
Parameters:
<state> ON | OFF
*RST: OFF
Example:
BUFF:STAT OFF
[SENSe<Sensor>:][POWer:][AVG:]FAST <state>
Enables or disables a fast unchopped Continuous Average measurement. The aver-
age count is enforced to 1. Any setting for average count is silently ignored. See also
"Calculating the measurement time" on page 93.
Parameters:
<state> *RST: OFF
Example: See Chapter 10, "Performing Measurement Tasks - Program-
ming Examples", on page 144.
Configuring the Measurement Modes