EasyManua.ls Logo

REPLIGEN XCell Lab XC-LAB-S - Table 6. A2 B Tubing ID;OD; Table 7. SAPA Components; Figure 7. SAPA Components Example

Default Icon
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XCell™ Lab Controller
User Guide
19
repligen.com XC-LAB-UG-V4
Table 6. A2B Tubing ID/OD
Device Tubing ID /OD Repligen Supplied Tube set?
XCell ATF® 1 1/8” / 1/4" Yes, part of device
XCell ATF® 2 1/4” / 3/8” Yes, supplied separately
XCell ATF® 4 3/8” / 5/8” Yes, supplied separately
Please reach out to your local Repligen representative for more information.
6.6 Supply Air Protection Assembly (SAPA)
Figure 7. SAPA components example
Table 7. SAPA components
Item Description
1 Manifold inlet Pressurized air from air supply
2 Filter unit
Filters air routing from air supply to manifold
Filter porosity: 0.3
3 Pressure regulator
Factory set air pressure supply regulator designed to
downregulate the supply air to a lower pressure required to
operate XCell ATF® 1, XCell ATF® 2 and XCell ATF® 4 (do not
adjust)
4 Pressure relief valve
Factory set air supply pressure relief valve set to relieve if
inlet pressure exceeds the specification noted above (do not
adjust)
5 Positive pressure supply manifold Pressurized air to XCell™ Lab Controller(s)
6 Manifold outlets Distributes air pressure to up to four XCell™ Lab Controllers
The SAPA normalizes air pressure from the lab utility line down to the required 15.6 psi. The
minimum utility air pressure requirement is 30 psi. A pressure relief valve provides safety in case of
the unlikely failure of the regulator (Figure 7)
. The regulator and relief valve are pre-set at the
factory; no modification is required by the end user. Installation should be performed or supervised
by an authorized Repligen service engineer.
Note: The SAPA must be installed with the air filter oriented downwards and plumbed.

Table of Contents