EasyManua.ls Logo

REPLIGEN XCell Lab XC-LAB-S - Settings Sub-Menu; Xcell ATF Device Configuration; Figure 22. Settings Sub-Menu Example; Figure 23. Xcell ATF Device Configuration Example

Default Icon
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XCell™ Lab Controller
User Guide
33
repligen.com XC-LAB-UG-V4
1
2
4
3
7.8.7 Settings sub-menu
Figure 22. Settings sub-menu example
1. Settings
2. XCell ATF® Device Configuration
3. Bioreactor Configuration
4. Default Configuration
5. General Configuration
6. Service/Performance
The settings sub-menu (Figure 22
) d
isplays the options available to customize the device, bioreactor,
controller, and software settings.
7
.8.8 XCell ATF® Device configuration
Figure 23. XCell ATF® Device configuration example
1. XCell ATF® Device size, format
2. Filter P/N (dropdown)
3. Default filter settings
4. Enable/disable P3 sensor
The XCell ATF® Device configuration sub-menu (Figure 23
) allo
ws selection of the XCell ATF®
Device(s) with a single click. Selection of a device opens default values and displays an image of the
chosen device. To run only one device with the D or D-P controller models, select NONE for the
other; device and the graphic will be removed from the device image appears blank. The ATF Main
screen will be updated as well.
O
nce an XCell ATF® Device selection is made, a filter part number can be selected from the drop-
down list. This enables the system to display the correct physical characteristics of the chosen filter
(shown below in grey), which is crucial for scale-up calculations.
Note: Uncheck the P3 box when not in use to avoid alarms due to lack of communication with the P3
sensor.
2 3 4 5 6
1

Table of Contents