Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev A BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc. 5-17
2. Date: Each event is date-stamped in a month/day/year format (for
example, 10/20/02).
3. Time: Each event is time-stamped in hour/minute/second format (for
example, 09:15:23).
4. Old value: For settings changes, the previously selected value is
displayed.
5. New value: For settings changes, the most recently selected value is
displayed.
Use the arrow keys to scroll through the log.
NOTE: Table 5-3 lists event IDs, and recommends corrective actions if
appropriate. When performing corrective actions:
Perform the corrective actions in the order listed (if applicable)
until the problem is resolved.
Before replacing a part, slave in a replacement part to verify that
it corrects the problem. If so, verify that the problem recurs with
the original part installed, then replace the part and confirm that
the problem is corrected.
Significant Event Log: Event IDs
Code Short text for code Comment / Corrective Action
1 ERR SOFTWARE Controller processor detected a software error.
If this persists and leads to multiple resets and the
system declaring a system error because too many
errors occurred in 24 hours, replace the controller PCB.
2 ERR PROGRAM CRC FAILURE Controller processor detected a program memory
checksum error.
If this persists replace the controller PCB.
3 ERR EXTERNAL RAM
FAILURE
Controller processor detected an external RAM memory
error.
If this persists replace the controller PCB.
4 ERR FIQ STACK OVERFLOW Controller processor detected a stack overflow.
If this persists and leads to multiple resets and the
system declaring a system error because too many
errors occurred in 24 hours, replace the controller PCB.
Table 5-3: Significant Event Log: Event IDs (Sheet 1 of 22)