RIGOL 2 Command System
2-306 MSO2000A/DS2000A Programming Guide
:TRIGger:SHOLd:TYPe
Syntax
:TRIGger:SHOLd:TYPe <type>
:TRIGger:SHOLd:TYPe?
Description
Set or query the hold type of setup/hold trigger.
Parameter
Name Type Range Default
<type> Discrete {SETup|HOLd|SETHOLd} SETup
Explanation
SETup: set the time (refer to the
:TRIGger:SHOLd:STIMe command) that the data stays stable and
constant before the clock edge appears.
HOLd: set the time (refer to the
:TRIGger:SHOLd:HTIMe command) that the data stays stable and
constant after the clock edge appears.
SETHOLd: set the time (refer to the
:TRIGger:SHOLd:STIMe and :TRIGger:SHOLd:HTIMe
commands) that the data stays stable and constant before and after the clock edge appears.
Return Format
The query returns SET, HOL or SETHOL.