RIGOL 2 Command System
2-314 MSO2000A/DS2000A Programming Guide
:TRIGger:SPI:SLOPe
Syntax
:TRIGger:SPI:SLOPe <slope>
:TRIGger:SPI:SLOPe?
Description
Set or query the edge type of the clock signal in SPI trigger.
Parameter
Name Type Range Default
<slope> Discrete {POSitive|NEGative} POSitive
Return Format
The query returns POS or NEG.
:TRIGger:SPI:TIMeout
Syntax
:TRIGger:SPI:TIMeout <time_value>
:TRIGger:SPI:TIMeout?
Description
Set or query the timeout time in SPI trigger when the trigger condition is Timeout. The unit is s.
Parameter
Name Type Range Default
<time_value> Real 100ns to 1s 1us
Return Format