43Chapter 3: Operation GX-3R Pro Operator’s Manual
3. If the error code is anything but 031, the instrument cannot be used. Contact RKI Instruments,
Inc. as soon as possible.
If the error code is 031, you may press and release POWER MODE to continue into
Measuring Mode if the instrument must be used temporarily.
CAUTION: There will be no datalogging function if you operate the instrument after a 031
system failure. Contact RKI Instruments, Inc. as soon as possible.
Responding to a Man Down Warning 1 and Warning 2
The Man Down Warning 1 and Warning 2 alarms occur after the WAR N IN G 1 T I M E and
WARNING 2 TIME, respectively, has passed since the last movement of the instrument. See
page 100 for instructions to change these values.
1. Follow your established procedure for a man down warning.
2. To silence the alarm and reset the Man Down clock, move the instrument or press and release
POWER MODE.
Responding to a Man Down Alarm
The Man Down Alarm alarms occur after the ALARM TIME has passed since the last
movement of the instrument. See page 100 for instructions to change this value.
1. Follow your established procedure for a man down alarm.
2. To silence the alarm and reset the Man Down clock, press and release POWER MODE or
AIR. Moving the instrument will not reset the alarm or Man Down clock.
Responding to a Panic Alarm
If the user is in a dangerous situation or feels that others must be alerted to any sort of problem,
forcefully tapping the instrument twice will initiate a panic alarm.
1. Press and release POWER MODE or AIR to silence and reset the alarm.