GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration94
6. See “ESCAPE” on page 98 to return to User Mode.
CAL TIME
The CAL TIME is the total time the instrument is exposed to calibration gas when a bump test
fails if AUTO CAL is set to ON. The bump test time is deducted from the calibration time. For
example, if the CAL TIME is set to 90 seconds and the GAS TIME is set to 30 seconds, if the
bump test fails, the GX-3R Pro will only be exposed to gas for an additional 60 seconds. The
available values are 90 seconds (factory setting), and 120 seconds.
1. While in the BUMP SETTING\BUMP PARAMETER menu, use AIR to place the cursor
next to CAL TIME.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUMP PARAMETER
menu.
5. See “ESCAPE” on page 96 to return to the BUMP SETTING menu.
6. See “ESCAPE” on page 98 to return to User Mode.
AUTO CAL
ON (factory setting): If a bump test fails, the unit automatically begins a calibration.
OFF: If a bump test fails, the unit does not automatically begin a calibration.
1. While in the BUMP SETTING\BUMP PARAMETER menu, use AIR to place the cursor
next to AUTO CAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUMP PARAMETER
menu.
5. See “ESCAPE” on page 96 to return to the BUMP SETTING menu.
6. See “ESCAPE” on page 98 to return to User Mode.