97Chapter 5: User Mode and Calibration GX-3R Pro Operator’s Manual
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUMP SETTING
menu.
5. See “ESCAPE” on page 98 to return to User Mode.
BUMP CHECK GAS
ALL GAS (factory setting): Bump test dates for all gases are used to determine if bump test is
due.
4 GAS: Bump test dates for only the standard 4 gases (combustible gas/O
2
/H
2
S/CO) are used to
determine if bump test is due.
1. While in the BUMP SETTING menu, use AIR to place the cursor next to BUMP CHECK
GAS.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUMP SETTING
menu.
5. See “ESCAPE” on page 98 to return to User Mode.
ESCAPE
1. While in the BUMP SETTING menu, use AIR to place the cursor next to ESCAPE.
2. Press and release POWER MODE. The instrument will return to User Mode.
3. See “Returning to Measuring Mode (START MEASURE)” on page 117 to return to
Measuring Mode.
MAINTE 7:49
BUMP EXPIRED
BUMP CHECK GAS
ESCAPE
BUMP SETTING
MAINTE 7:49
BUMP EXPIRED
BUMP CHECK GAS
ESCAPE
BUMP SETTING