GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration96
BUMP INTERVAL
This setting defines the amount of time between bump tests. The time can be set in 1 day
increments. The minimum setting is 0 days and the maximum setting is 30 days. The factory
setting is 30 days.
1. While in the BUMP SETTING menu, use AIR to place the cursor next to BUMP
INTERVAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUMP SETTING
menu.
5. See “ESCAPE” on page 98 to return to User Mode.
BUMP EXPIRED
This item defines what indication is given during start up when a bump test is due and BUMP
REMINDER is set to ON.
CONFIRM TO USE (factory setting): The GX-3R Pro gives an indication at start up if a bump
test is past due. Press and release AIR to continue without bump testing or POWER MODE to
perform a bump test.
CANNOT USE: The GX-3R Pro gives an indication at start up if a bump test is past due. Press
and release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no
effect. A successful bump test must be performed in order to use the instrument.
NO EFFECT
: The GX-3R Pro gives an indication at startup if a bump test is past due. If desired,
press POWER MODE to perform a bump test but it is not necessary to acknowledge the bump
test due indication. The warm-up sequence will continue on its own.
1. While in the BUMP SETTING menu, use AIR to place the cursor next to BUMP
EXPIRED.
2. Press and release POWER MODE. The current setting flashes.
MAINTE 7:49
BUMP REMINDER
BUMP INTERVAL
BUMP EXPIRED
BUMP SETTING
MAINTE 7:49
BUMP INTERVAL
BUMP EXPIRED
BUMP CHECK GAS
BUMP SETTING