Control Testing
61
English
Display of results
Disposal of used items after control testing
Patient Self-Testers
Discard the used test strip with your regular household waste. For more details, refer to the pack-
age insert.
Healthcare Professionals
Refer to the package insert.
Dispose of used consumables in line with the disposal policy of your hospital, institute or
medical practice.
When measurement is finished, the result is
displayed.
Check whether the displayed result falls
within the acceptable range.
Target values which should be obtained with
control solutions can either be found on the
labels or on the package inserts for the test
strips or control solutions. If the value is out-
side the range, repeat the control test. If the
second result is again outside this range,
please contact your local customer support
and service centre (see page 81).