Product Specifications
79
English
10 Product Specifications
Operating conditions and technical data
Sample material
Temperature range for measure-
ments with patient samples
Dependent on test parameter:
• For cholesterol: 18-35 °C
• For glucose: 18-35 °C
• For triglycerides: 18-30 °C
• For lactate: 15-35 °C
Temperature range for measure-
ments with control solutions
Dependent on test parameter:
• For cholesterol: 18-30 °C
• For glucose: 18-32 °C
• For triglycerides: 18-30 °C
• For lactate: 15-35 °C
Relative humidity 10 to 85%
Measurement range Blood glucose: 20–600 mg/dL (1.1–33.3 mmol/L)
Cholesterol: 150–300 mg/dL (3.88–7.76 mmol/L)
Triglycerides: 70–600 mg/dL (0.80–6.86 mmol/L)
Lactate: 0.8–21.7 mmol/L (blood value), 0.7–26 mmol/L
(plasma value)
Memory 100 test results, optionally with date, time, and additional infor-
mation per test parameter
Battery operation 4 x AAA 1.5 V alkaline manganese batteries
Number of tests
per set of batteries
More than 1,000 tests (with new batteries)
Safety class III
Dimensions 154 x 81 x 30 mm
Weight Approx. 140 g
Sample type Fresh capillary blood (for detailed information see package
insert)
Sample size A hanging drop of blood
Interactions Refer to test strip package insert