Cleaning
71
English
8 Cleaning
A clean optical measuring system is a basic prerequisite for determining precise results. There-
fore clean the meter regularly and immediately after it becomes dirty. Always power off the meter
before cleaning it.
Use only the following items for cleaning:
Ordinary lint-free cotton buds
Ordinary lint-free tissues
Ordinary disinfecting tissues
Recommended cleaning/disinfecting solutions
Mild soap suds as well as 70 % ethanol or isopropyl alcohol are suitable for cleaning.
Healthcare Professionals: When used in the professional sector (e.g. doctor’s office), a mixture
consisting of 1-propanol (400 mg/g), 2-propanol (200 mg/g) and glutaraldehyde (1.0 mg/g)
1
is
recommended.
Cleaning the exterior (meter housing)
Ensure the meter is powered off.
Wipe the outside of the meter with a lightly moistened, lint-free cotton cloth.
1. Sold in some countries under the name “Bacillol Plus.”
NOTICE
Damage to the instrument through moisture
Ensure that no liquid enters the meter. If moisture enters the meter, it
may cause malfunction of the instrument.
Do not use any disinfectant sprays or tissues or cotton buds which are
dripping wet as the liquid may enter the meter and damage it.