Troubleshooting
76
Error/Display Description/Solution
Error: Test strip is
unusable
The test strip has already been used or is dirty.
Solution
Remove and discard the strip. Repeat the test using a new test
strip.
Error: Temperature
The meter is too hot or cold, or the ambient temperature does not
fall within the acceptable range for the selected test
Solution
Move the meter to an environment which has an appropriate
temperature and repeat the test after several minutes. Do not
artifically heat or cool the meter by any means.
Error: Flap open
(after meter has been
powered on)
The measurement chamber flap is not closed properly.
Solution
Close the measurement chamber flap.
Error: Flap open (during
measurement)
The measurement chamber flap is not closed properly.
Solution
Close the measurement chamber flap.