Control and Proficiency Testing • 5
119
7 Apply gentle pressure to the reservoir to transfer
the entire contents of the dropper to the bottle.
Make sure that the dropper does not come into
contact with the dried control plasma.
8 Close the bottle again.
9 Make sure the dropper is at hand for the next steps
in the liquid quality control test.
10 Swirl the bottle using a circular motion to com-
pletely dissolve all of the control plasma inside. Do
not shake the bottle or turn it on its side. Doing
so can cause components in the control plasma to
stick to the sides of the bottle. Please refer to the
control solution package insert.
The control solution is now ready to be applied to the test
strip.
The control solutions may be reconstituted (mixed) after
removal from the refrigerator. For further information refer
to the respective package insert(s).