Meter Setup • 3
77
Patient ID
In the default setting, input of Patient IDs is set to No. This
means each test is simply assigned a consecutive
number. However, you can require that a Patient ID be
entered or make it optional.
1 From the Main Menu, touch Setup to open the
meter settings.
2 From the Setup Menu, touch ID Setup.
3 From the ID Setup menu, touch Patient.
You may select from the following options:
■ No (tests will be assigned numbers automatically)
■ Optional (automatic numbering or
list/scan/manual input)
■ Required (list/scan/manual input)
4 Touch the button with the setting of choice. Your
selection is now highlighted.
5 Touch to save this setting, or touch to exit
this menu without saving any changes.
The settings for the option No are now completed. For the
options Optional and Required, continue by selecting the
input format.
If you want to create a list of Patient IDs from which you
can select a patient for testing, additional software (a
data management system) and the Handheld Base Unit
are required (see page 139).
ID Setup
09:15 am
05/29/2015
Patient
Operator
Admin.
Setup Menu
09:15 am
05/29/2015
QC Settings
ID Setup
Diagnostics
Options
Screen
Main Menu
09:15 am
05/29/2015
Control Test
Review Results
Setup
Patient T
est
Patient ID
09:15 am
05/29/2015
No
Optional
Required