General Product Specifications • 10
159
10 General Product Specifications
10.1 Technical data
Sample material
Temperature range +12 °C to +32 °C (54 °F to 90 °F)
Relative humidity 10 to 85% (no condensation)
Maximum altitude 4300 m (14,000 feet)
Position Place the meter on a level, vibration-free surface or hold it so it
is roughly horizontal.
Measuring range Refer to respective test strip package insert
Memory 2000 patient and 500 QC results with date and time
120 code chip records (60 strip codes and 60 control codes)
Operator lists up to 5000 Operator IDs with corresponding 2nd
ID (e.g., Operator Name)
Patient lists up to 4000 Patient IDs with corresponding 2nd and
3rd Patient IDs (e.g., name, date of birth)
Interface Touchscreen and barcode scanner
Battery operation Universal Battery Pack for the CoaguChek Pro II meter
Power connection Power supply adapter: Input: 100-240 V / 50-60 Hz / 350-150 mA
Output: 12 V DC / 1.25 A
Number of tests with a fully charged
battery pack
approx. 60 tests (PT/INR)
at least 20 tests (aPTT/measuring time 5 minutes)
Safety class III
Automatic power-off Programmable 1 … 60 minutes
Dimensions 187 x 97 x 43 mm
Weight 280 g (without battery pack)
Sample type Refer to test strip package insert
Sample size Refer to test strip package insert
Interactions Refer to test strip package insert