Operator lists After powering the meter on, an Operator login may be
configured. There are four possible configurations for an
Operator login:
■ Operator is activated, but there is no list available:
The Operator can either log in via the onscreen
keypad or a barcode scan. The use of the meter is
not restricted to a predefined Operator group, so
Operator IDs are entered via the onscreen keypad
or read from the barcode and stored with test
results for informational purposes only.
■ Operator is activated, there is no list available and
the “List” feature is set to “hidden” (only possible
with a DMS). Setting the "List" feature to "hidden"
with the DMS automatically blocks the display of
the meter's onscreen keypad as well.
The Operator can only log in via a barcode scan as
there is no onscreen keypad. The use of the meter
is not restricted to a predefined Operator group, so
Operator IDs are read from the barcode and stored
with test results for informational purposes only.
Meters which are managed by a DMS may receive an
Operator list.
■ Operator is activated, a list is available and dis-
played on the login screen:
The Operator selects the ID from the displayed list.
If so configured by the Administrator, the Operator
may also be required to enter a password to log in.
■ Operator is activated, a list is available but set to
“hidden”:
The Operator can login via a barcode scan. If so
configured by the Administrator, the Operator may
also be required to enter a password to log in.
This first configuration is also available without a DMS.
When working with operator lists: The use of the meter