EasyManua.ls Logo

Roche CoaguCheck Pro II - Operator Lists

Roche CoaguCheck Pro II
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142
Operator lists After powering the meter on, an Operator login may be
configured. There are four possible configurations for an
Operator login:
Operator is activated, but there is no list available:
The Operator can either log in via the onscreen
keypad or a barcode scan. The use of the meter is
not restricted to a predefined Operator group, so
Operator IDs are entered via the onscreen keypad
or read from the barcode and stored with test
results for informational purposes only.
Operator is activated, there is no list available and
the “List” feature is set to “hidden” (only possible
with a DMS). Setting the "List" feature to "hidden"
with the DMS automatically blocks the display of
the meter's onscreen keypad as well.
The Operator can only log in via a barcode scan as
there is no onscreen keypad. The use of the meter
is not restricted to a predefined Operator group, so
Operator IDs are read from the barcode and stored
with test results for informational purposes only.
Meters which are managed by a DMS may receive an
Operator list.
Operator is activated, a list is available and dis-
played on the login screen:
The Operator selects the ID from the displayed list.
If so configured by the Administrator, the Operator
may also be required to enter a password to log in.
Operator is activated, a list is available but set to
“hidden”:
The Operator can login via a barcode scan. If so
configured by the Administrator, the Operator may
also be required to enter a password to log in.
This first configuration is also available without a DMS.
When working with operator lists: The use of the meter
is restricted to the operators on this list.

Table of Contents

Related product manuals