Introduction • 1
27
Operating conditions To ensure that the meter functions properly, please
observe the following guidelines:
Only use the meter at a room temperature between
12 °C and 32 °C (54 °F and 90 °F).
Only use the meter at a relative humidity between
10% and 85% (no condensation).
When operating the meter using the power adapter,
use only a voltage of 100 V to 240 V (± 10%) , 50/
60 Hz.
When testing, place the meter on a level, vibration-
free surface, or hold it so it is roughly horizontal.
Quality control The meter has a number of built-in quality-control
functions:
A check of the electronic components and func-
tions every time the meter is powered on.
A check of the test strip temperature while a test is
in progress.
A check of the expiration date and lot information
on the test strip based on the code chip data.
An onboard quality control within every single test
strip.
Roche Diagnostics offers liquid quality controls for the
CoaguChek Pro II system. These controls are designed to
assist you in meeting regulatory compliance require-
ments at your facility.