Extended Functionalities • 7
141
Computer (Setup option) For initial connection to a DMS, the ability to communi-
cate within a network has to be set up as follows:
In the Setup Menu, the Connection option must be
set to Computer (see page 63).
The Handheld Base Unit must be configured cor-
rectly. (For details please refer to the manual of the
Handheld Base Unit and the Technical Note stored
on the Handheld Base Unit itself.)
Place the meter on a Handheld Base Unit that is
connected to the network. The meter will automat-
ically be recognized by the DMS and may now be
assigned to a site and/or unit managed by the DMS
(if your DMS has that functionality).
Depending on DMS functionality available, the
DMS may transfer settings and lists to the meter as
well as enable test results to be transferred from the
meter to the DMS.