158
Meter reset A meter reset should only be performed if all other reme-
dies have failed.
1 Place the meter on a level surface.
2 Press the On/Off button for at least 12 seconds.
■ The meter powers off and on again.
■ The Roche logo is displayed.
If the Roche logo does not appear within 60 sec-
onds, place the meter in the Handheld Base Unit for
a minimum of 15 minutes to recharge the battery.
■ The meter performs a system check.
■ The screens for entering the date and the time
appear.
3 Enter the date and time. After you have entered the
correct information, confirm each screen with .
■ Unless all lockouts are disabled (see chapter 3.5),
the meter will now be in QC Lockout due to the
manually entered date/time.
4 Synchronize the meter’s date/time with the date/
time of your facility via the Handheld Base Unit or,
if working wirelessly, wait at least ten minutes for
the next WLAN synchronization before performing
any further tests.
Even if your configuration does not require it, we
recommend always performing a QC test after a meter
reset.