System Administrator (Admin.) In the default setting, the meter is not protected with an
Admin. ID, and all setup options are accessible to every
user. If you set up an Admin. ID, the following setup areas
are automatically reserved only for the system adminis-
trator/supervisor (i.e., the person who knows the
password).
■ Screen: Result Units, Result Confirmation, Date/
Time
■ Options: Connection (to a computer or a printer)
■ ID Setup (the entire area)
■ QC Settings (the entire area)
■ Diagnostics: Software, Hardware, Memory,
Wireless Setup
If an Admin. ID has not been set up yet:
1 From the Main Menu, touch Setup to open the
meter settings.
2 From the Setup Menu, touch ID Setup.
3 From the ID Setup menu , touch Admin.
When you enter an Admin. ID, this ID will have to be
entered from this point forward before any of the set-
tings above can be changed. The Admin. ID must also
be entered before you can delete or change the Admin.
ID itself. If you forget the Admin. ID, contact your Roche
Diagnostics representative.
ID Setup
09:15 am
05/29/2015
Patient
Operator
Admin.
Setup Menu
09:15 am
05/29/2015
QC Settings
ID Setup
Diagnostics
Options
Screen
Main Menu
09:15 am
05/29/2015
Control Test
Review Results
Setup
Patient T
est