112
Accepting or rejecting a test
result
If this function is enabled, you may choose to accept or
reject a test result. When the result is displayed, touch
Reject or
Accept
If you reject a result, you must enter a comment with an
explanation.
If you reject the result, this test result is no longer dis-
played. However, the test entry is stored.
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op: Operator 1
Par: PT
Measurement Result OK?
21
%Q
05/29/2015 09:14 am
AcceptReject
Comment Required
A comment is required if you
reject a result. Please enter a
comment.
I-009
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
Smith, Henr
y
Op: Operator 1
Par: PT
Will Repeat T
est
Rejected
05/29/2015 09:14 am