Testing a Blood Sample • 4
111
When interpreting results, refer to the detailed informa-
tion on limitations and interferences included in the limi-
tations section of the test strip package insert.
If a “C” is displayed along with the result:
This may occur if the hematocrit value is very low or due
to erroneous blood collection (e.g., wet hands). Repeat
the test. If using capillary blood, make sure that the
patient's hands are dry. If the message persists, perform
a hematocrit check.
If a “
*
” is displayed along with the result:
This may occur if the test result is out of normal range.
Touch the
*
icon to display the Out Of Normal Range
information screen.
Test
09:15 am
05/29/2015
P-ID: PID05
Code: 208
1.6
INR
Müller, David
Op: Schulz
Par: PT
05/29/2015 09:14 am
C
Test
09:15 am
05/29/2015
P-ID: PID05
Code: 208
1.6
INR
Müller, David
Op: Schulz
Par: PT
05/29/2015 09:14 am
C