110
You hear a beep tone when you have applied enough
blood (provided the Beeper is enabled). The blood drop
symbol disappears and the test starts.
The meter performs an automatic quality control test on
the test strip before it displays the test result. “QC”
appears in the display.
Following a successful outcome of the quality control
test, a checkmark appears after “QC.”
The result is displayed in the unit you chose when setting
up the meter. It is automatically saved.
For information on the valid measuring ranges for
CoaguChek PT test strips and for CoaguChek aPTT test
strips refer to the package insert of the respective test
strip.
Results that are above or below the measuring range are
indicated by the symbols > (above) or < (below).
Do not add more blood. To avoid error messages do not
touch the test strip until the result is displayed.
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
Smith, Henr
y
Op: Operator 1
Par: PT
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
Smith, Henr
y
Op: Operator 1
Par: PT
QC
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op: Operator 1
Par: PT
21
%Q
05/29/2015 09:14 am
Test
09:15 am
05/29/2015
P-ID: PID01
Code: 184
Smith, Henr
y
Op: Operator 1
Par: PT
05/29/2015 09:14 am
> 8.0INR