122
If you are using a new test strip lot and have not inserted
the test strip code chip yet, you must do so now. Other-
wise you cannot perform a quality control test.
As with the test strips, a quality control code chip is also
provided with the control solutions. This chip informs the
meter about the acceptable ranges of results for that lot
of controls. The information on the code chip is retained
in the memory so you can use the same control solutions
at any time.
12 Select the code stored for your current control
solution or touch New Code to use a new control
solution.
When entering or selecting a code manually, make sure it
matches the code printed on the bottle of the control
solution.
Alternatively, the control solution code can also be
entered using the built-in barcode scanner. Touch
Scan and scan the barcode on the label from a distance
of approx. 10 cm (4 inches).
The meter beeps once the barcode has been read suc-
cessfully. The barcode information appears in the code
field. The scanner turns off after 10 seconds, if a barcode
is not scanned.
QC Test
09:15 am
05/29/2015
184
Op: Operator 1
QC Test 09:15 am
05/29/2015
QC 070
QC 060
QC 050
New Code
QC 048
Scan