150
Follow the instructions given below to clean and disinfect
the meter. Failure to follow these instructions may cause
malfunction of the meter.
Observe the disinfection guidelines of your
institution.
Use gloves.
NOTICE
Instrument damage due to moisture
Ensure that no liquid enters the meter. If moisture enters
the meter, it may cause malfunction of the instrument.
Make sure the meter is turned off.
Do not spray anything onto the meter housing.
Do not spray into the test strip guide.
Do not immerse the meter in liquid.
Do not use cloths or swabs/buds that are saturated.
Squeeze off excess solution or blot on a dry paper
towel to remove any excess solution before wiping
the surface of the meter.